验证硬件 » SAIMD » SAIMD-1 (64 MHz)
SAIMD-1-64

SAIMD-1 (64 MHz)

Standard Active Implantable Device for Systems Verification at 64 MHz

The SAIMD-1 (64 MHz) is a verification device for evaluations according to ISO/TS 10974 at 1.5 T. 

SAIMD-1 (64 MHz).

 

   

Wire Material

Stainless steel

Insulator Material

Polyolefin

Liquid Compatibility

Compatible with sugar- and oil-based TSM.
Not compatible with TSM based on Triton or diethylene glycol monobutyl ether (DGBE).

Dimensions

Wire length: 200 ± 1 mm
Tips length: 10 ± 0.3 mm
Conductor diameter: 1.5 ± 0.1 mm
Thickness insulator: 0.5 ± 0.1 mm

Cerificates & Target Values

Certificates with transfer function as well as SAR values vs. distance from tip for the piX and ELIT1.5 phantoms, respectively, are available for the following tissue simulating media (TSM):

Compatible Phantoms

piX, ELIT1.5, ASTM2009