验证硬件 » SAIMD

SAIMD

Standard Active Implantable Medical Devices for RF Safety Evaluation at MRI Frequencies

The SAIMD package from ZMT includes verification and validation devices for instrumentations and procedures according to ISO TS/10974 for 1.5 T and/or 3.0 T test systems.

  Length
Frequency Application

SAIMD-1 (64 MHz)

200 mm

64 MHz

system verification for RF heating evaluation
according to Clause 8 of ISO/TS 10974

SAIMD-1 (128 MHz)

100 mm

128 MHz

system verification for RF heating evaluation
according to Clause 8 of ISO/TS 10974

SAIMD-U

600 mm

64 and 128 MHz

system verification and validation for RF model
generation, as well as heating and induced
voltage evaluation, according to Clause 8 and
Clause 15 of ISO/TS 10974, respectively,
with DUAL-CYL-TFD